Region

Applications

Extraction of Opiates in Human Serum Using Strata-X PRO and a Kinetex 2.6 µm Biphenyl LC Column

Sample Preparation Details

Application Detail (App ID: 25490)

Part No:00B-4622-Y0
Dimensions:50 x 3 mm ID
Elution Type:Gradient
Elution A:0.1%Formic acid in Water
Elution B:0.1%Formic acid in Methanol
Gradient Profile:
Step No. Time (min) Pct A Pct B
1 0 90 5
2 4 5 95
3 5.5 5 95
4 5.51 95 5
5 7 95 5

Flow Rate:0.5 mL/min
Col. Temp: 45 °C
Detection: LC/MS/MS @ (0°C)
Note:Human Serum Analyte diluted with initial mobile phase

Analyte Details

Analyte Detail not available

1
Morphine
search

2
Oxymorphone
search

3
Hydromorphone
search

4
Naloxone
search

5
6-MAM
search

6
Codeine
search

7
Oxycodone
search

8
Hydrocodone
search

9
Norfentanyl
search

10
Fentanyl
search